Mechanical Specification

Acrylic double walled canopy with front loading & six elbow operated port holes

Sliding acrylic baby tray

Facility to take X-ray

Three drawer storage cabinet & shelf to keep essential things for baby

Mounted on heavy duty castors

Stainless steel blower sub assembly

Inlets for oxygen, IV tubings, etc.

Humidity tray & humidity indicator

External head up/down positioning facility
Micro Computer Temperature Controller

Provides thermal regulation with highest degree of accuracy

User friendly feather touch control

Skin control, air control & manual control

For skin control: 34 to 38°C

For air control : Room temp. to 39°C

Resolution : 0.1°C

Accuracy: ± 0.2°C within set temp.

Safety alarms along with high temp. Cutoff.

Power source: 230 volts ± 10%, 50 Hz.
SKIN / AIR MANUAL ( Micro Computer Temperature Controllers ) |
 1? bright skin temp. display
 1? bright air temp. display
 % of heater output display
|
 Control mode display on LCD
 Set temp. display on LCD
|
| Manual Mode |
 Time programmable
 Heater output settable
 Over temp. cutoff.
|
Timer / APGAR Timer Baby attending alarm Auto audio alarm mute Mute time programmable Safety Alarms Messages on LCD |
Skin Mode |
 Skin high 0.5° C
 Skin low 0.5° C
 Skin over range (>38° C)
|
 Skin under range (<34° C)
 Skin probe failure
|
| Air Mode |
Air high 1.5° C
Air low 3.0° C |
Air over range
Air probe failure
|
|
Safety high temp. cutoff at |
 Skin temp. >38° C
 Air temp. >39° C
|
| Heater & Power failure alarm |